Instructions for use
SMARTtorque LUX S619 L - 1.008.1641
SMARTtorque Mini LUX S615 L - 1.008.1643
SMARTtorque S619 C - 1.008.1642
Always be on the safe side.
Zobrazit stránku 0
1 2 3 4 5 6 ... 24 25

Shrnutí obsahu

Strany 1 - Instructions for use

Instructions for useSMARTtorque LUX S619 L - 1.008.1641SMARTtorque Mini LUX S615 L - 1.008.1643SMARTtorque S619 C - 1.008.1642Always be on the safe si

Strany 2

4.3 Checking the pressuresCAUTIONCompressed air connection on devices.Dirty and moist compressed air causes premature bearing wear.▶ Ensure that the

Strany 3 - Table of contents

5 OperationNoteAt the beginning of each workday, the water-conducting systems should berinsed for at least 2 min. without the instrument being attac

Strany 4 - Target group

WARNINGUse of unauthorised cutters or grinders.Injury to the patient or damage to the medical device.▶ Observe the instructions for use and use the cu

Strany 5 - 2.2 Safety instructions

6 Troubleshooting6.1 Exchanging the O-rings on the coupling on the supply hoseCAUTIONHazard from improper care of the O-rings.Malfunctions or comp

Strany 6

7 Reprocessing methods according to ISO 176647.1 Preparations at the site of useWARNINGHazard from nonsterile products.There is a risk of infectio

Strany 7 - Proper use:

▶ In order to prevent negative effects on the medical device, make sure that theinterior and the exterior of the medical device are dry after completi

Strany 8

7.3.1 Manual external disinfectionKaVo recommends the following products based on material compatibility.▪ Mikrozid AF made by Schülke & Mayr (l

Strany 9 - 4 Start up and shut down

7.4 DryingManual Drying▶ Blow off the outside and inside with compressed air until water drops are nolonger visible.Automatic DryingThe drying proce

Strany 10 - 4.4 Checking the O-rings

▶ Remove the cutter or grinder, place the INTRA spray nipple tip in the openingand spray.▶ Subsequently treat with the specified care agents and syste

Strany 11 - 5 Operation

NoteClose the chuck service mode.Option 1: Place the dental handpieces in the QUATTROcare PLUS 2124 A,close the front door and start theservice proced

Strany 12

Distributed by:KaVo Dental GmbHBismarckring 39D-88400 BiberachPhone +49 7351 56-0Fax +49 7351 56-1488Manufacturer:Kaltenbach & Voigt GmbHBismarckr

Strany 13 - 6 Troubleshooting

▶ Use according to the manufacturer's Instructions for Use.7.8 Storage▶ Reprocessed products should be stored protected from dust with minimum

Strany 14 - 7.2 Cleaning

8 ToolsAvailable from dental suppliers.Material summary Mat. no.Replacement rotor Standard X.XXX.XXXXWrench Standard X.XXX.XXXXInstrument stand 2151

Strany 15 - 7.3 Disinfection

9 Warranty terms and conditionsThe following warranty conditions apply to this KaVo medical device:The KaVo service offices listed in the appendix p

Strany 16

PolandKaVo Polska Sp.z.o.o.UL. Pomorska 251 p.405PL - 92-213 LodzPhone: +48 42 675 75 36Fax: +48 42 675 75 35Email: [email protected]: www.kavo.p

Strany 18

1.008.1927 · kb · 20111124 - 03 · en

Strany 19 - 7.7 Sterilisation

Table of contents1 User instructions ...

Strany 20 - 7.8 Storage

1 User instructionsDear UserCongratulations on purchasing this KaVo quality product. By following the instruc‐tions below you will be able to work s

Strany 21 - 8 Tools

2 Safety2.1 Description of safety instructionsWarning symbolStructureDANGERThe introduction describes the type and source of the hazard.This secti

Strany 22

WARNINGHazard from contraindication.If the soft tissue in the oral cavity is injured, the compressed air may enable sep‐tic substances to enter the ti

Strany 23

3 Product descriptionSMARTtorque LUX S619 L (Mat. no. 1.008.1641)SMARTtorque Mini LUX S615 L (Mat. no. 1.008.1643)SMARTtorque S619 C (Mat. no. 1.008

Strany 24

3.2 Technical Specifications SURGtorque S459 LMinimal drive pressure 2.1 bar (30 psi)Drive pressure 2.1 to 4.2 bar (30 to 61 psi)Air consumption 40

Strany 25

4 Start up and shut downWARNINGHazard from non-sterile products.Infection hazard for care provider and patient.▶ Before first use and after each use